Bremsstrahlung radiation
Also called: braking radiation, continuous X-ray spectrum
Definition
Bremsstrahlung (German for braking) radiation is produced inside the X-ray tube when a high-speed projectile electron from the cathode passes close to the positively charged nucleus of a tungsten target atom, slows down, and changes direction. The kinetic energy lost by the decelerating electron is emitted directly as an X-ray photon. Because electrons pass at varying distances from target nuclei and lose anywhere from a tiny fraction to all of their kinetic energy, bremsstrahlung produces a continuous spectrum of photon energies up to the selected kVp, accounting for the majority of diagnostic X-ray photons.
How Bremsstrahlung radiation Is Tested on the Registry Exam
ARRT questions distinguish bremsstrahlung (nuclear electrostatic deceleration producing a continuous energy spectrum) from characteristic radiation (inner-shell electron transition at discrete energies).
ARRT Radiography Exam Overview (2026)
Bremsstrahlung radiation is part of the core curriculum for the ARRT Radiography Certification (R.T.(R)) exam.
- Questions
- 230 (200 scored)
- Time Limit
- 3 hours 50 minutes
- Passing Score
- 75 (1-99)
- Test Provider
- Pearson VUE
Official Radiography Content Domains
Textbook & Clinical References
- Bushong, S. C. Radiologic Science for Technologists: Physics, Biology, and Protection.
Related Radiography Physics, Image Production and Radiation Safety Terms
View all 22 terms →Characteristic radiation
Characteristic radiation is produced in the X-ray tube target when a projectile electron has enough kinetic energy to eject an inner-shell (K-shell) electron from a target atom, leaving a vacancy that is filled by an outer-shell electron dropping down. The emitted X-ray photon has a discrete energy exactly equal to the difference in binding energies between the two shells. For a tungsten target, K-shell electrons have a binding energy of 69.5 keV, so useful K-characteristic X-rays are produced only when the tube is operated at 70 kVp or higher.
Read definition & exam tips →
Grid ratio and grid cutoff
Grid ratio is the height of the lead strips in an anti-scatter radiographic grid divided by the width of the radiolucent interspace material between them (r = h/D). Higher grid ratios (such as 12:1 versus 8:1) clean up more scattered radiation and improve image contrast, but require a higher mAs (Bucky factor) and increase patient dose. Higher-ratio grids also have a narrower positioning latitude, making them more prone to grid cutoff, an undesirable absorption of primary beam photons caused by off-center, off-level, off-focus, or upside-down grid alignment.
Read definition & exam tips →
Magnification factor (MF)
The magnification factor (MF) quantifies how much larger an anatomic structure appears on the radiographic image compared with its true object size. It is calculated either as image size divided by object size, or as source-to-image receptor distance (SID) divided by source-to-object distance (SOD), where SOD equals SID minus object-to-image receptor distance (OID). Minimizing magnification and improving spatial resolution requires using a long SID (such as 72 inches for a chest radiograph) and keeping OID as close to zero as possible.
Read definition & exam tips →
Source-to-image receptor distance (SID)
Source-to-image receptor distance (SID) is the linear distance from the X-ray tube focal spot to the image receptor, typically standardized at 40 inches (102 cm) for most table and upright exams and 72 inches (183 cm) for chest and cervical spine radiography. Increasing SID reduces beam divergence across the anatomy, decreasing size distortion (magnification) and focal-spot blur to improve spatial resolution. Because beam intensity drops with distance according to the inverse square law, increasing SID requires a compensating increase in mAs (using the direct square law) to maintain receptor exposure.
Read definition & exam tips →
Object-to-image receptor distance (OID)
Object-to-image receptor distance (OID) is the distance between the anatomic part being imaged and the image receptor. Of all geometric factors, OID has the most direct impact on magnification and focal-spot blur: as OID increases, the diverging X-rays spread further apart before striking the detector, enlarging the image and reducing sharpness. When anatomy creates an unavoidable OID (such as a lateral cervical spine), technologists compensate by increasing SID to 72 inches; the resulting air gap also allows scattered photons to miss the detector.
Read definition & exam tips →
Automatic exposure control (AEC)
Automatic exposure control (AEC) is a system of radiation detectors (typically three ionization chambers positioned between the patient and the image receptor) that measures the radiation reaching the receptor and automatically terminates the X-ray exposure once a preset air kerma threshold is reached. The technologist still selects the kVp, mA, active chamber configuration, and backup timer, while the AEC controls the exposure time. Accurate centering of the anatomy of interest directly over the active chamber is essential; if the part is off-center and primary beam strikes an uncovered chamber, the exposure terminates prematurely.
Read definition & exam tips →