Dynamic range
Definition
Dynamic range is the ratio between the largest and smallest echo amplitudes a system can process and display, expressed in decibels. A wide dynamic range shows many shades of gray, a softer-looking, lower-contrast image, while a narrow dynamic range shows fewer shades and higher contrast. Compression shrinks that ratio before display: apply 50 dB of compression to a 90 dB signal and 40 dB reaches the screen.
How Dynamic range Is Tested on the Registry Exam
SPI questions ask how widening or narrowing dynamic range changes the number of gray shades, and how much compression lowers the final dynamic range.
ARDMS SPI Exam Overview (2026)
Dynamic range is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Dynamic range
Review these related video lessons from Prepry's registry review courses.
Textbook & Clinical References
- Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.
Related Ultrasound Physics and Instrumentation Terms
View all 27 terms →Harmonic imaging
Harmonic imaging forms the image from harmonic frequencies generated as sound distorts while traveling through tissue, rather than from the reflected fundamental frequency. The second harmonic used is exactly twice the transmitted frequency. Because harmonic energy builds up gradually with depth and is weakest in the off-axis side lobes, tissue harmonic imaging reduces near-field reverberation clutter and lobe artifact compared with fundamental imaging, improving contrast resolution in technically difficult patients.
Read definition & exam tips →
Anechoic, hypoechoic, isoechoic, and hyperechoic
Echogenicity describes how bright a tissue or lesion appears on grayscale B-mode ultrasound relative to adjacent reference tissue. An anechoic structure produces zero internal echoes and displays completely black, which is the classic appearance of simple fluid such as a simple cyst, the urinary bladder, or a patent blood vessel lumen. Hypoechoic structures return weaker echoes and look darker gray than surrounding parenchyma, isoechoic structures match the brightness of the adjacent tissue, and hyperechoic structures return stronger echoes and appear brighter white (such as fat, fibrous tissue, calcifications, or air).
Read definition & exam tips →
Mechanical index and thermal index
Mechanical index (MI) and thermal index (TI) are the two real-time on-screen output display standard (ODS) bioeffect indices that help sonographers practice ALARA. Mechanical index estimates the likelihood of non-thermal mechanical bioeffects, specifically cavitation (microbubble expansion and collapse), and is proportional to peak rarefactional pressure divided by the square root of frequency, meaning lower frequency and higher acoustic output raise MI. Thermal index estimates the potential temperature rise in tissue (TIS for soft tissue, TIB for bone at the focus, and TIC for cranial bone at the surface), where a TI of 1.0 indicates a potential 1 degree Celsius rise.
Read definition & exam tips →
Coded excitation
Coded excitation is a transmit processing technique created in the pulser that sends a long, mathematically encoded sequence of pulses (such as a frequency-modulated chirp or Golay code) into the body and then mathematically compresses the returning echo on receive into a very short, high-amplitude spike. Because the energy is spread across a longer transmit burst rather than a single high-voltage spike, coded excitation increases signal-to-noise ratio, penetration, and contrast resolution while preserving axial resolution and keeping peak intensity (and mechanical index) within safe bioeffect limits.
Read definition & exam tips →
Acoustic impedance
Acoustic impedance is a tissue's resistance to the passage of sound, measured in Rayls, and it equals the tissue's density multiplied by the propagation speed of sound through it. At a boundary struck straight on, reflection depends on how much the impedance of the two tissues differs: identical impedances transmit all the sound, and a larger mismatch reflects more of it back to the transducer. This is why a soft tissue to bone or air interface reflects so strongly.
Read definition & exam tips →
Propagation speed
Propagation speed is how fast sound travels through a medium, and it depends only on the medium's stiffness and density, not on the transducer. Speed rises with stiffness and falls with density, so sound travels fastest in bone, around 3,500 m/s, and slowest in air, while soft tissue is assumed to average 1,540 m/s. The machine uses that assumed value, together with echo return time, to calculate the depth of a reflector.
Read definition & exam tips →