Mechanical index and thermal index
Also called: mechanical index ultrasound, MI, thermal index, TI, TIB, TIS, TIC, cavitation
Definition
Mechanical index (MI) and thermal index (TI) are the two real-time on-screen output display standard (ODS) bioeffect indices that help sonographers practice ALARA. Mechanical index estimates the likelihood of non-thermal mechanical bioeffects, specifically cavitation (microbubble expansion and collapse), and is proportional to peak rarefactional pressure divided by the square root of frequency, meaning lower frequency and higher acoustic output raise MI. Thermal index estimates the potential temperature rise in tissue (TIS for soft tissue, TIB for bone at the focus, and TIC for cranial bone at the surface), where a TI of 1.0 indicates a potential 1 degree Celsius rise.
How Mechanical index and thermal index Is Tested on the Registry Exam
SPI Safety questions test that raising acoustic output power raises both MI and TI, lowering transmit frequency raises MI (increasing cavitation risk), and obstetric scans require keeping TI below 1.0 and minimizing pulsed spectral Doppler exposure time.
ARDMS SPI Exam Overview (2026)
Mechanical index and thermal index is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Mechanical index and thermal index
Review these related video lessons from Prepry's registry review courses.
Abdominal Transplants and Postoperative Assessment - Part 1
Explore sonographic evaluation of liver, kidney, and pancreas transplants including preoperative vascular mapping and postoperative monitoring. Master Doppler techniques for assessing hepatic artery patency, resistive index values, and identifying complications like thrombosis. ARDMS, ARRT, and CCI registry exam preparation.
Abdominal Imaging Optimization
Master the principles of ultrasound physics including frequency, wavelength, and attenuation to optimize abdominal imaging. Learn how to select the appropriate curvilinear or linear transducer based on patient habitus and target depth. ARDMS, ARRT, and CCI registry exam preparation.
Imaging Artifacts - Part 1
Explore the physics of ultrasound artifacts, including propagation speed errors and equipment assumptions. Learn to differentiate between anatomic artifacts like posterior shadowing and enhancement versus equipment malfunctions. ARDMS, ARRT, and CCI registry exam preparation.
Imaging Artifacts - Part 2
Explore equipment-related artifacts including electrical interference, transducer malfunction, and vertical dropout. Master the identification of twinkle artifact for renal calculi and understand the physics behind spectral Doppler aliasing and the Nyquist limit. ARDMS, ARRT, and CCI registry exam preparation.
Textbook & Clinical References
- AIUM Official Statement: As Low As Reasonably Achievable (ALARA) Principle.
Related Ultrasound Physics and Instrumentation Terms
View all 27 terms →Coded excitation
Coded excitation is a transmit processing technique created in the pulser that sends a long, mathematically encoded sequence of pulses (such as a frequency-modulated chirp or Golay code) into the body and then mathematically compresses the returning echo on receive into a very short, high-amplitude spike. Because the energy is spread across a longer transmit burst rather than a single high-voltage spike, coded excitation increases signal-to-noise ratio, penetration, and contrast resolution while preserving axial resolution and keeping peak intensity (and mechanical index) within safe bioeffect limits.
Read definition & exam tips →
Acoustic impedance
Acoustic impedance is a tissue's resistance to the passage of sound, measured in Rayls, and it equals the tissue's density multiplied by the propagation speed of sound through it. At a boundary struck straight on, reflection depends on how much the impedance of the two tissues differs: identical impedances transmit all the sound, and a larger mismatch reflects more of it back to the transducer. This is why a soft tissue to bone or air interface reflects so strongly.
Read definition & exam tips →
Propagation speed
Propagation speed is how fast sound travels through a medium, and it depends only on the medium's stiffness and density, not on the transducer. Speed rises with stiffness and falls with density, so sound travels fastest in bone, around 3,500 m/s, and slowest in air, while soft tissue is assumed to average 1,540 m/s. The machine uses that assumed value, together with echo return time, to calculate the depth of a reflector.
Read definition & exam tips →
Attenuation
Attenuation is the progressive weakening of an ultrasound beam as it travels through tissue, from absorption, scattering, and reflection. In soft tissue the attenuation coefficient in dB/cm equals half the frequency in MHz, and total attenuation in decibels equals that coefficient multiplied by the path length in cm. Because the coefficient rises with frequency, higher-frequency transducers and deeper structures both lose more signal, which is why raising frequency trades penetration for resolution.
Read definition & exam tips →
Axial resolution
Axial resolution is the ability to display two structures lying one behind the other, along the beam's path, as two separate echoes. It is determined entirely by spatial pulse length, since axial resolution equals spatial pulse length divided by two, so shorter pulses give better, smaller-number resolution. Raising frequency or damping the transducer more heavily shortens the pulse and improves it. The depth of the reflector itself has no effect on axial resolution.
Read definition & exam tips →
Lateral resolution
Lateral resolution is the ability to display two structures lying side by side, perpendicular to the beam, as two separate echoes, and it is set by the width of the ultrasound beam at a given depth. It is best at the focus, where the beam is narrowest, and worsens in the near and far fields where the beam is wider. Raising frequency and properly placing the focal zone at the depth of interest both narrow the beam and improve lateral resolution.
Read definition & exam tips →