Harmonic imaging
Definition
Harmonic imaging forms the image from harmonic frequencies generated as sound distorts while traveling through tissue, rather than from the reflected fundamental frequency. The second harmonic used is exactly twice the transmitted frequency. Because harmonic energy builds up gradually with depth and is weakest in the off-axis side lobes, tissue harmonic imaging reduces near-field reverberation clutter and lobe artifact compared with fundamental imaging, improving contrast resolution in technically difficult patients.
How Harmonic imaging Is Tested on the Registry Exam
SPI questions ask why harmonic imaging reduces reverberation and lobe artifact by relying on energy generated deeper in tissue and mainly along the main beam.
ARDMS SPI Exam Overview (2026)
Harmonic imaging is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Harmonic imaging
Review these related video lessons from Prepry's registry review courses.
Imaging Artifacts - Part 2
Explore equipment-related artifacts including electrical interference, transducer malfunction, and vertical dropout. Master the identification of twinkle artifact for renal calculi and understand the physics behind spectral Doppler aliasing and the Nyquist limit. ARDMS, ARRT, and CCI registry exam preparation.
Harmonics and Contrast Enhanced Ultrasound
Understand the principles of non-linear sound propagation and the generation of harmonic frequencies within biological tissues. Explore how tissue harmonic imaging improves lateral resolution and reduces reverberation artifacts to enhance diagnostic clarity. ARDMS, ARRT, and CCI registry exam preparation.
Textbook & Clinical References
- Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.
Related Ultrasound Physics and Instrumentation Terms
View all 27 terms →Anechoic, hypoechoic, isoechoic, and hyperechoic
Echogenicity describes how bright a tissue or lesion appears on grayscale B-mode ultrasound relative to adjacent reference tissue. An anechoic structure produces zero internal echoes and displays completely black, which is the classic appearance of simple fluid such as a simple cyst, the urinary bladder, or a patent blood vessel lumen. Hypoechoic structures return weaker echoes and look darker gray than surrounding parenchyma, isoechoic structures match the brightness of the adjacent tissue, and hyperechoic structures return stronger echoes and appear brighter white (such as fat, fibrous tissue, calcifications, or air).
Read definition & exam tips →
Mechanical index and thermal index
Mechanical index (MI) and thermal index (TI) are the two real-time on-screen output display standard (ODS) bioeffect indices that help sonographers practice ALARA. Mechanical index estimates the likelihood of non-thermal mechanical bioeffects, specifically cavitation (microbubble expansion and collapse), and is proportional to peak rarefactional pressure divided by the square root of frequency, meaning lower frequency and higher acoustic output raise MI. Thermal index estimates the potential temperature rise in tissue (TIS for soft tissue, TIB for bone at the focus, and TIC for cranial bone at the surface), where a TI of 1.0 indicates a potential 1 degree Celsius rise.
Read definition & exam tips →
Coded excitation
Coded excitation is a transmit processing technique created in the pulser that sends a long, mathematically encoded sequence of pulses (such as a frequency-modulated chirp or Golay code) into the body and then mathematically compresses the returning echo on receive into a very short, high-amplitude spike. Because the energy is spread across a longer transmit burst rather than a single high-voltage spike, coded excitation increases signal-to-noise ratio, penetration, and contrast resolution while preserving axial resolution and keeping peak intensity (and mechanical index) within safe bioeffect limits.
Read definition & exam tips →
Acoustic impedance
Acoustic impedance is a tissue's resistance to the passage of sound, measured in Rayls, and it equals the tissue's density multiplied by the propagation speed of sound through it. At a boundary struck straight on, reflection depends on how much the impedance of the two tissues differs: identical impedances transmit all the sound, and a larger mismatch reflects more of it back to the transducer. This is why a soft tissue to bone or air interface reflects so strongly.
Read definition & exam tips →
Propagation speed
Propagation speed is how fast sound travels through a medium, and it depends only on the medium's stiffness and density, not on the transducer. Speed rises with stiffness and falls with density, so sound travels fastest in bone, around 3,500 m/s, and slowest in air, while soft tissue is assumed to average 1,540 m/s. The machine uses that assumed value, together with echo return time, to calculate the depth of a reflector.
Read definition & exam tips →
Attenuation
Attenuation is the progressive weakening of an ultrasound beam as it travels through tissue, from absorption, scattering, and reflection. In soft tissue the attenuation coefficient in dB/cm equals half the frequency in MHz, and total attenuation in decibels equals that coefficient multiplied by the path length in cm. Because the coefficient rises with frequency, higher-frequency transducers and deeper structures both lose more signal, which is why raising frequency trades penetration for resolution.
Read definition & exam tips →