Reverberation
Definition
Reverberation is an artifact caused by sound bouncing back and forth repeatedly between two strong, closely spaced reflectors, or between the transducer face and a reflector, before finally returning. Each bounce adds travel time, so the machine displays a series of equally spaced parallel echoes at increasing depth, fading in brightness with each successive line as energy is lost. It is a recognized sign of free air, such as pneumoperitoneum next to the diaphragm.
How Reverberation Is Tested on the Registry Exam
SPI and abdomen questions ask you to identify reverberation from its ladder of equally spaced, fading parallel echoes, including as a sign of free intraperitoneal air.
ARDMS SPI Exam Overview (2026)
Reverberation is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Reverberation
Review these related video lessons from Prepry's registry review courses.
Harmonics and Contrast Enhanced Ultrasound
Understand the principles of non-linear sound propagation and the generation of harmonic frequencies within biological tissues. Explore how tissue harmonic imaging improves lateral resolution and reduces reverberation artifacts to enhance diagnostic clarity. ARDMS, ARRT, and CCI registry exam preparation.
Unit 24 US Physics, Understanding Artifacts and Optimizing Image Quality Pt 2
Explore ultrasound artifacts in OB/GYN imaging including reverberation, comet tail, and ring down. Understand how propagation speed errors and equipment assumptions impact image quality and diagnostic accuracy. Differentiate between pathology and anatomic artifacts for ARDMS, ARRT, and CCI registry exam preparation.
1.4 Artifact Recognition
Identify common 2D and Doppler artifacts, understand why they occur, and select practical corrections that improve diagnostic confidence and image quality.
Textbook & Clinical References
- Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.
Related Artifacts and Image Optimization Terms
View all 11 terms →Comet tail artifact
Comet tail artifact is a form of reverberation seen as a single, tapering, bright line of echoes trailing behind a small, strongly reflective structure, such as a metal clip, cholesterol crystals, or gas. The reflectors are so closely spaced that the individual reverberation echoes merge into one narrowing, tail-like streak instead of staying as separate parallel lines. It is the classic sign of adenomyomatosis of the gallbladder wall and of pneumobilia after ERCP.
Read definition & exam tips →
Ring-down artifact
Ring-down artifact is a continuous, solid vertical band of echoes extending posteriorly from a small collection of gas bubbles, rather than the separate lines of typical reverberation. It is caused by fluid trapped between adjacent gas bubbles resonating at the transmitted frequency and re-emitting sound of their own, producing a sustained signal instead of discrete echoes. It is a recognized sign of gas, such as in the gastrointestinal tract, and differs from comet tail artifact, which tapers and is tied to metal or crystals.
Read definition & exam tips →
Acoustic shadowing
Acoustic shadowing is an artifact appearing as a dark band deep to a structure that strongly reflects or attenuates sound, such as a gallstone, calcification, or bone. Most of the beam stops at that structure, so the tissue underneath receives almost nothing, sends back almost nothing, and paints dark on the screen. It is the opposite of posterior acoustic enhancement, which brightens tissue behind a weak attenuator.
Read definition & exam tips →
Posterior acoustic enhancement
Posterior acoustic enhancement is an artifact appearing as an unusually bright band deep to a structure, such as a simple cyst, that attenuates sound less than the surrounding tissue. Because time gain compensation assumes typical soft tissue attenuation and the fluid absorbs less sound than that, more energy actually reaches tissue beyond it, making that deeper tissue look falsely brighter. It is a helpful sign that a well-defined anechoic lesion, like a breast, splenic, or ganglion cyst, is fluid-filled rather than solid.
Read definition & exam tips →
Mirror image artifact
Mirror image artifact is a duplicate of a real structure displayed deeper than it truly is, produced when sound reflects off a strong, curved specular reflector, classically the diaphragm, continues to the real structure, and bounces back by the same path before returning to the transducer. Because the machine assumes a straight path, it places the copy along the beam but at the greater depth matching the extra travel time. A related form, crosstalk, mirrors a spectral waveform across the baseline.
Read definition & exam tips →
Refraction artifact
Refraction artifact occurs when the beam bends at a boundary between two media with different propagation speeds, struck at an oblique angle, following Snell's law. This can duplicate a structure so it appears just to the side of its true position, or produce edge shadowing, a thin, clean shadow at the lateral margins of a curved structure like a cyst, gallbladder, or carotid artery, where the beam grazes the surface and bends away instead of returning.
Read definition & exam tips →