Ultrasound Artifacts (ARDMS SPI)

Mirror image artifact

Definition

Mirror image artifact is a duplicate of a real structure displayed deeper than it truly is, produced when sound reflects off a strong, curved specular reflector, classically the diaphragm, continues to the real structure, and bounces back by the same path before returning to the transducer. Because the machine assumes a straight path, it places the copy along the beam but at the greater depth matching the extra travel time. A related form, crosstalk, mirrors a spectral waveform across the baseline.

How Mirror image artifact Is Tested on the Registry Exam

SPI and abdomen questions use liver lesions mirrored above the diaphragm as the classic example, and always place the artifact deeper than the true structure.

ARDMS SPI Exam Overview (2026)

Mirror image artifact is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Mirror image artifact

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.

Related Artifacts and Image Optimization Terms

View all 11 terms →

Refraction artifact

Refraction artifact occurs when the beam bends at a boundary between two media with different propagation speeds, struck at an oblique angle, following Snell's law. This can duplicate a structure so it appears just to the side of its true position, or produce edge shadowing, a thin, clean shadow at the lateral margins of a curved structure like a cyst, gallbladder, or carotid artery, where the beam grazes the surface and bends away instead of returning.

Read definition & exam tips →

Side lobe artifact

Side lobe artifact is a false-echo artifact caused by the weaker secondary beams, called side lobes, that a transducer emits off-axis in addition to its main beam. When a side lobe strikes a strong reflector outside the main beam, such as bowel gas, the returning echo is displayed as though it came from directly along the main axis, placing false echoes inside a structure that should be anechoic, like the gallbladder or urinary bladder.

Read definition & exam tips →

Speed error artifact

Speed error artifact is the mispositioning of a reflector that results from the machine always assuming sound travels at a constant 1,540 m/s to calculate depth, even when the true speed in a specific tissue, such as fat or a breast implant, is different. If the actual speed is slower than assumed, the extra travel time makes the machine place the reflector too deep; if it is faster, the reflector displays too shallow.

Read definition & exam tips →

Spatial compounding

Spatial compounding combines several real-time frames of the same tissue, acquired by steering the beam to different angles, into one averaged composite image. Because random speckle and angle-dependent artifacts like edge shadowing appear differently at each angle, averaging cancels much of that noise while true tissue boundaries reinforce, improving border definition. The trade-off is a lower frame rate, since the system must acquire and combine more frames for each displayed image, reducing temporal resolution.

Read definition & exam tips →

Anisotropy

Anisotropy is an angle-dependent ultrasound artifact seen in tightly organized fibrillar structures, most notably tendons, ligaments, nerves, and muscle, whose echogenicity changes dramatically depending on the angle of incidence of the ultrasound beam. When the beam strikes a healthy tendon perpendicularly (at a 90 degree angle), its tightly packed collagen fibers act as strong specular reflectors and appear brightly hyperechoic. Tilting the transducer even 2 to 5 degrees off perpendicular causes the reflected beam to miss the transducer face, making the normal tendon appear falsely hypoechoic or anechoic and mimicking tendinosis or a partial-thickness tear.

Read definition & exam tips →

Reverberation

Reverberation is an artifact caused by sound bouncing back and forth repeatedly between two strong, closely spaced reflectors, or between the transducer face and a reflector, before finally returning. Each bounce adds travel time, so the machine displays a series of equally spaced parallel echoes at increasing depth, fading in brightness with each successive line as energy is lost. It is a recognized sign of free air, such as pneumoperitoneum next to the diaphragm.

Read definition & exam tips →