Posterior acoustic enhancement
Definition
Posterior acoustic enhancement is an artifact appearing as an unusually bright band deep to a structure, such as a simple cyst, that attenuates sound less than the surrounding tissue. Because time gain compensation assumes typical soft tissue attenuation and the fluid absorbs less sound than that, more energy actually reaches tissue beyond it, making that deeper tissue look falsely brighter. It is a helpful sign that a well-defined anechoic lesion, like a breast, splenic, or ganglion cyst, is fluid-filled rather than solid.
How Posterior acoustic enhancement Is Tested on the Registry Exam
Abdomen and breast questions use posterior enhancement behind a well-defined anechoic lesion as supporting evidence that it is a simple, fluid-filled cyst rather than solid.
ARDMS SPI Exam Overview (2026)
Posterior acoustic enhancement is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Posterior acoustic enhancement
Review these related video lessons from Prepry's registry review courses.
Urinary System Pathology
Explore sonographic features of renal cell carcinoma, simple cysts, and angiomyolipomas. Learn to differentiate benign and malignant urinary masses using Doppler ultrasound and transducer selection for retroperitoneal imaging. ARDMS, ARRT, and CCI registry exam preparation.
Pathology of the Liver and Spleen - Part 1: Liver Pathology
Explore sonographic assessment of splenic pathologies including splenomegaly, infarction, and hemangiomas. Understand how lab values like CBC and LDH correlate with splenic health and identify wedge-shaped hypoechoic lesions. ARDMS, ARRT, and CCI registry exam preparation.
Textbook & Clinical References
- Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound. (p. 359)
Related Artifacts and Image Optimization Terms
View all 11 terms →Mirror image artifact
Mirror image artifact is a duplicate of a real structure displayed deeper than it truly is, produced when sound reflects off a strong, curved specular reflector, classically the diaphragm, continues to the real structure, and bounces back by the same path before returning to the transducer. Because the machine assumes a straight path, it places the copy along the beam but at the greater depth matching the extra travel time. A related form, crosstalk, mirrors a spectral waveform across the baseline.
Read definition & exam tips →
Refraction artifact
Refraction artifact occurs when the beam bends at a boundary between two media with different propagation speeds, struck at an oblique angle, following Snell's law. This can duplicate a structure so it appears just to the side of its true position, or produce edge shadowing, a thin, clean shadow at the lateral margins of a curved structure like a cyst, gallbladder, or carotid artery, where the beam grazes the surface and bends away instead of returning.
Read definition & exam tips →
Side lobe artifact
Side lobe artifact is a false-echo artifact caused by the weaker secondary beams, called side lobes, that a transducer emits off-axis in addition to its main beam. When a side lobe strikes a strong reflector outside the main beam, such as bowel gas, the returning echo is displayed as though it came from directly along the main axis, placing false echoes inside a structure that should be anechoic, like the gallbladder or urinary bladder.
Read definition & exam tips →
Speed error artifact
Speed error artifact is the mispositioning of a reflector that results from the machine always assuming sound travels at a constant 1,540 m/s to calculate depth, even when the true speed in a specific tissue, such as fat or a breast implant, is different. If the actual speed is slower than assumed, the extra travel time makes the machine place the reflector too deep; if it is faster, the reflector displays too shallow.
Read definition & exam tips →
Spatial compounding
Spatial compounding combines several real-time frames of the same tissue, acquired by steering the beam to different angles, into one averaged composite image. Because random speckle and angle-dependent artifacts like edge shadowing appear differently at each angle, averaging cancels much of that noise while true tissue boundaries reinforce, improving border definition. The trade-off is a lower frame rate, since the system must acquire and combine more frames for each displayed image, reducing temporal resolution.
Read definition & exam tips →
Anisotropy
Anisotropy is an angle-dependent ultrasound artifact seen in tightly organized fibrillar structures, most notably tendons, ligaments, nerves, and muscle, whose echogenicity changes dramatically depending on the angle of incidence of the ultrasound beam. When the beam strikes a healthy tendon perpendicularly (at a 90 degree angle), its tightly packed collagen fibers act as strong specular reflectors and appear brightly hyperechoic. Tilting the transducer even 2 to 5 degrees off perpendicular causes the reflected beam to miss the transducer face, making the normal tendon appear falsely hypoechoic or anechoic and mimicking tendinosis or a partial-thickness tear.
Read definition & exam tips →