Ultrasound Artifacts (ARDMS SPI)

Anisotropy

Also called: anisotropy ultrasound, anisotropic artifact

Definition

Anisotropy is an angle-dependent ultrasound artifact seen in tightly organized fibrillar structures, most notably tendons, ligaments, nerves, and muscle, whose echogenicity changes dramatically depending on the angle of incidence of the ultrasound beam. When the beam strikes a healthy tendon perpendicularly (at a 90 degree angle), its tightly packed collagen fibers act as strong specular reflectors and appear brightly hyperechoic. Tilting the transducer even 2 to 5 degrees off perpendicular causes the reflected beam to miss the transducer face, making the normal tendon appear falsely hypoechoic or anechoic and mimicking tendinosis or a partial-thickness tear.

How Anisotropy Is Tested on the Registry Exam

SPI, pediatric, and musculoskeletal ultrasound questions ask how to distinguish true tendon pathology from anisotropy by heel-toeing or rocking the transducer back to a 90 degree angle of incidence.

ARDMS SPI Exam Overview (2026)

Anisotropy is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Textbook & Clinical References

Related Artifacts and Image Optimization Terms

View all 11 terms →

Reverberation

Reverberation is an artifact caused by sound bouncing back and forth repeatedly between two strong, closely spaced reflectors, or between the transducer face and a reflector, before finally returning. Each bounce adds travel time, so the machine displays a series of equally spaced parallel echoes at increasing depth, fading in brightness with each successive line as energy is lost. It is a recognized sign of free air, such as pneumoperitoneum next to the diaphragm.

Read definition & exam tips →

Comet tail artifact

Comet tail artifact is a form of reverberation seen as a single, tapering, bright line of echoes trailing behind a small, strongly reflective structure, such as a metal clip, cholesterol crystals, or gas. The reflectors are so closely spaced that the individual reverberation echoes merge into one narrowing, tail-like streak instead of staying as separate parallel lines. It is the classic sign of adenomyomatosis of the gallbladder wall and of pneumobilia after ERCP.

Read definition & exam tips →

Ring-down artifact

Ring-down artifact is a continuous, solid vertical band of echoes extending posteriorly from a small collection of gas bubbles, rather than the separate lines of typical reverberation. It is caused by fluid trapped between adjacent gas bubbles resonating at the transmitted frequency and re-emitting sound of their own, producing a sustained signal instead of discrete echoes. It is a recognized sign of gas, such as in the gastrointestinal tract, and differs from comet tail artifact, which tapers and is tied to metal or crystals.

Read definition & exam tips →

Acoustic shadowing

Acoustic shadowing is an artifact appearing as a dark band deep to a structure that strongly reflects or attenuates sound, such as a gallstone, calcification, or bone. Most of the beam stops at that structure, so the tissue underneath receives almost nothing, sends back almost nothing, and paints dark on the screen. It is the opposite of posterior acoustic enhancement, which brightens tissue behind a weak attenuator.

Read definition & exam tips →

Posterior acoustic enhancement

Posterior acoustic enhancement is an artifact appearing as an unusually bright band deep to a structure, such as a simple cyst, that attenuates sound less than the surrounding tissue. Because time gain compensation assumes typical soft tissue attenuation and the fluid absorbs less sound than that, more energy actually reaches tissue beyond it, making that deeper tissue look falsely brighter. It is a helpful sign that a well-defined anechoic lesion, like a breast, splenic, or ganglion cyst, is fluid-filled rather than solid.

Read definition & exam tips →

Mirror image artifact

Mirror image artifact is a duplicate of a real structure displayed deeper than it truly is, produced when sound reflects off a strong, curved specular reflector, classically the diaphragm, continues to the real structure, and bounces back by the same path before returning to the transducer. Because the machine assumes a straight path, it places the copy along the beam but at the greater depth matching the extra travel time. A related form, crosstalk, mirrors a spectral waveform across the baseline.

Read definition & exam tips →