Ultrasound Artifacts (ARDMS SPI)

Spatial compounding

Definition

Spatial compounding combines several real-time frames of the same tissue, acquired by steering the beam to different angles, into one averaged composite image. Because random speckle and angle-dependent artifacts like edge shadowing appear differently at each angle, averaging cancels much of that noise while true tissue boundaries reinforce, improving border definition. The trade-off is a lower frame rate, since the system must acquire and combine more frames for each displayed image, reducing temporal resolution.

How Spatial compounding Is Tested on the Registry Exam

SPI questions test that spatial compounding reduces speckle and angle-dependent artifact like edge shadowing, at the cost of lower frame rate and temporal resolution.

ARDMS SPI Exam Overview (2026)

Spatial compounding is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.

Related Artifacts and Image Optimization Terms

View all 11 terms →

Anisotropy

Anisotropy is an angle-dependent ultrasound artifact seen in tightly organized fibrillar structures, most notably tendons, ligaments, nerves, and muscle, whose echogenicity changes dramatically depending on the angle of incidence of the ultrasound beam. When the beam strikes a healthy tendon perpendicularly (at a 90 degree angle), its tightly packed collagen fibers act as strong specular reflectors and appear brightly hyperechoic. Tilting the transducer even 2 to 5 degrees off perpendicular causes the reflected beam to miss the transducer face, making the normal tendon appear falsely hypoechoic or anechoic and mimicking tendinosis or a partial-thickness tear.

Read definition & exam tips →

Reverberation

Reverberation is an artifact caused by sound bouncing back and forth repeatedly between two strong, closely spaced reflectors, or between the transducer face and a reflector, before finally returning. Each bounce adds travel time, so the machine displays a series of equally spaced parallel echoes at increasing depth, fading in brightness with each successive line as energy is lost. It is a recognized sign of free air, such as pneumoperitoneum next to the diaphragm.

Read definition & exam tips →

Comet tail artifact

Comet tail artifact is a form of reverberation seen as a single, tapering, bright line of echoes trailing behind a small, strongly reflective structure, such as a metal clip, cholesterol crystals, or gas. The reflectors are so closely spaced that the individual reverberation echoes merge into one narrowing, tail-like streak instead of staying as separate parallel lines. It is the classic sign of adenomyomatosis of the gallbladder wall and of pneumobilia after ERCP.

Read definition & exam tips →

Ring-down artifact

Ring-down artifact is a continuous, solid vertical band of echoes extending posteriorly from a small collection of gas bubbles, rather than the separate lines of typical reverberation. It is caused by fluid trapped between adjacent gas bubbles resonating at the transmitted frequency and re-emitting sound of their own, producing a sustained signal instead of discrete echoes. It is a recognized sign of gas, such as in the gastrointestinal tract, and differs from comet tail artifact, which tapers and is tied to metal or crystals.

Read definition & exam tips →

Acoustic shadowing

Acoustic shadowing is an artifact appearing as a dark band deep to a structure that strongly reflects or attenuates sound, such as a gallstone, calcification, or bone. Most of the beam stops at that structure, so the tissue underneath receives almost nothing, sends back almost nothing, and paints dark on the screen. It is the opposite of posterior acoustic enhancement, which brightens tissue behind a weak attenuator.

Read definition & exam tips →

Posterior acoustic enhancement

Posterior acoustic enhancement is an artifact appearing as an unusually bright band deep to a structure, such as a simple cyst, that attenuates sound less than the surrounding tissue. Because time gain compensation assumes typical soft tissue attenuation and the fluid absorbs less sound than that, more energy actually reaches tissue beyond it, making that deeper tissue look falsely brighter. It is a helpful sign that a well-defined anechoic lesion, like a breast, splenic, or ganglion cyst, is fluid-filled rather than solid.

Read definition & exam tips →