Radiography & X-Ray Physics (ARRT)

Photoelectric effect

Also called: photoelectric absorption, photoelectron

Definition

The photoelectric effect is an interaction in which an incident X-ray photon transfers all of its energy to a tightly bound inner-shell (typically K-shell) electron of a tissue atom, completely absorbing the photon and ejecting the electron as a photoelectron. Another electron then drops into the inner-shell vacancy, releasing low-energy characteristic radiation. The likelihood of photoelectric absorption rises sharply with the third power of the absorber's atomic number (Z^3) and falls rapidly as photon energy (kVp) increases, which is why dense bone and barium or iodine contrast agents appear bright white on radiographs.

How Photoelectric effect Is Tested on the Registry Exam

ARRT questions contrast the photoelectric effect (inner-shell electron, total absorption, creates differential subject contrast, increases patient dose) with Compton scattering.

ARRT Radiography Exam Overview (2026)

Photoelectric effect is part of the core curriculum for the ARRT Radiography Certification (R.T.(R)) exam.

Full Radiography Study Guide →
Questions
230 (200 scored)
Time Limit
3 hours 50 minutes
Passing Score
75 (1-99)
Test Provider
Pearson VUE

Official Radiography Content Domains

Patient Care16.5%
Safety25%
Image Production25.5%
Procedures33%

Textbook & Clinical References

  • Bushong, S. C. Radiologic Science for Technologists: Physics, Biology, and Protection.

Related Radiography Physics, Image Production and Radiation Safety Terms

View all 22 terms →

Bremsstrahlung radiation

Bremsstrahlung (German for braking) radiation is produced inside the X-ray tube when a high-speed projectile electron from the cathode passes close to the positively charged nucleus of a tungsten target atom, slows down, and changes direction. The kinetic energy lost by the decelerating electron is emitted directly as an X-ray photon. Because electrons pass at varying distances from target nuclei and lose anywhere from a tiny fraction to all of their kinetic energy, bremsstrahlung produces a continuous spectrum of photon energies up to the selected kVp, accounting for the majority of diagnostic X-ray photons.

Read definition & exam tips →

Characteristic radiation

Characteristic radiation is produced in the X-ray tube target when a projectile electron has enough kinetic energy to eject an inner-shell (K-shell) electron from a target atom, leaving a vacancy that is filled by an outer-shell electron dropping down. The emitted X-ray photon has a discrete energy exactly equal to the difference in binding energies between the two shells. For a tungsten target, K-shell electrons have a binding energy of 69.5 keV, so useful K-characteristic X-rays are produced only when the tube is operated at 70 kVp or higher.

Read definition & exam tips →

Grid ratio and grid cutoff

Grid ratio is the height of the lead strips in an anti-scatter radiographic grid divided by the width of the radiolucent interspace material between them (r = h/D). Higher grid ratios (such as 12:1 versus 8:1) clean up more scattered radiation and improve image contrast, but require a higher mAs (Bucky factor) and increase patient dose. Higher-ratio grids also have a narrower positioning latitude, making them more prone to grid cutoff, an undesirable absorption of primary beam photons caused by off-center, off-level, off-focus, or upside-down grid alignment.

Read definition & exam tips →

Magnification factor (MF)

The magnification factor (MF) quantifies how much larger an anatomic structure appears on the radiographic image compared with its true object size. It is calculated either as image size divided by object size, or as source-to-image receptor distance (SID) divided by source-to-object distance (SOD), where SOD equals SID minus object-to-image receptor distance (OID). Minimizing magnification and improving spatial resolution requires using a long SID (such as 72 inches for a chest radiograph) and keeping OID as close to zero as possible.

Read definition & exam tips →

Source-to-image receptor distance (SID)

Source-to-image receptor distance (SID) is the linear distance from the X-ray tube focal spot to the image receptor, typically standardized at 40 inches (102 cm) for most table and upright exams and 72 inches (183 cm) for chest and cervical spine radiography. Increasing SID reduces beam divergence across the anatomy, decreasing size distortion (magnification) and focal-spot blur to improve spatial resolution. Because beam intensity drops with distance according to the inverse square law, increasing SID requires a compensating increase in mAs (using the direct square law) to maintain receptor exposure.

Read definition & exam tips →

Object-to-image receptor distance (OID)

Object-to-image receptor distance (OID) is the distance between the anatomic part being imaged and the image receptor. Of all geometric factors, OID has the most direct impact on magnification and focal-spot blur: as OID increases, the diverging X-rays spread further apart before striking the detector, enlarging the image and reducing sharpness. When anatomy creates an unavoidable OID (such as a lateral cervical spine), technologists compensate by increasing SID to 72 inches; the resulting air gap also allows scattered photons to miss the detector.

Read definition & exam tips →