X-ray circuit (primary, secondary & filament circuits)
Also called: xray circuit, secondary circuit, primary circuit, filament circuit, step-up transformer, step-down transformer
Definition
The radiographic X-ray imaging system divides electrical power into three interconnected sections: the primary (low-voltage) circuit, the secondary (high-voltage) circuit, and the filament circuit. The primary circuit contains the main power switch, circuit breakers, autotransformer (kVp selector via self-induction), timer circuit (AEC or electronic timer), and the primary side of the high-voltage step-up transformer. The secondary circuit contains the secondary coil of the step-up transformer (which increases voltage from volts to kilovolts with a turns ratio of 500:1 to 1,000:1) and the solid-state rectifier bank (four diodes for full-wave rectification converting AC to pulsed DC before reaching the X-ray tube). The filament circuit uses a variable rheostat (mA selector) and a step-down transformer to reduce voltage to ~10 V while increasing current to 3 to 5 A to heat the tungsten cathode filament for thermionic emission.
How X-ray circuit (primary, secondary & filament circuits) Is Tested on the Registry Exam
ARRT Image Production questions frequently ask which components sit in the primary circuit (autotransformer, exposure timer, primary step-up coil), the secondary circuit (secondary step-up coil, rectifier bridge, X-ray tube), and the filament circuit (mA selector, step-down transformer, focal spot selector).
ARRT Radiography Exam Overview (2026)
X-ray circuit (primary, secondary & filament circuits) is part of the core curriculum for the ARRT Radiography Certification (R.T.(R)) exam.
- Questions
- 230 (200 scored)
- Time Limit
- 3 hours 50 minutes
- Passing Score
- 75 (1-99)
- Test Provider
- Pearson VUE
Official Radiography Content Domains
Textbook & Clinical References
- Bushong SC. Radiologic Science for Technologists: Physics, Biology, and Protection. 12th ed. Elsevier; 2021.
Related Radiography Physics, Image Production and Radiation Safety Terms
View all 31 terms →Spatial resolution in radiography & digital imaging
Spatial resolution in radiography (traditionally called recorded detail or sharpness) is the imaging system's ability to distinguish two adjacent small anatomic structures as separate entities, measured in line pairs per millimeter (lp/mm) or modulation transfer function (MTF). In digital radiography (DR and CR), intrinsic detector spatial resolution is governed primarily by detector element (del) size and pixel pitch (smaller pixel size and smaller pixel pitch increase spatial resolution). Geometric factors that maximize spatial resolution and minimize penumbra (geometric unsharpness) include using a small effective focal spot size, increasing source-to-image receptor distance (SID), and minimizing object-to-image receptor distance (OID).
Read definition & exam tips →
Clements-Nakayama method (modified axiolateral hip projection)
The Clements-Nakayama modification is a unilateral trauma axiolateral projection of the hip performed when a patient with a suspected femoral neck or intertrochanteric fracture cannot abduct, flex, or rotate either leg (such as when bilateral leg splints or traction are present, precluding the Danelius-Miller method). With the patient supine, the grid image receptor is placed in a vertical cradle beside the affected hip and tilted 15 degrees posteriorly, while the central ray (CR) is angled 15 to 20 degrees posteriorly (mediolaterally) and directed perpendicular to the coronal plane of the femoral neck.
Read definition & exam tips →
Nasal bones X-ray positioning & routine projections (Lateral & Waters)
The routine radiographic series for the nasal bones consists of bilateral horizontal-beam or recumbent lateral projections (right and left lateral for comparison, examining the side closest to the image receptor) and a parietoacanthial (Waters) projection (with the orbitomeatal line at 37 degrees to the image receptor so the mentomeatal line is perpendicular). On the lateral nasal bone projection, the interpupillary line (IPL) is positioned perpendicular to the image receptor, the infraorbitomeatal line (IOML) is parallel to the transverse axis of the IR, and the perpendicular central ray (CR) enters 1/2 inch (1.25 cm) inferior to the nasion with tight close collimation and low kVp (50 to 60 kVp, non-grid detail technique) to visualize subtle nondisplaced nasal bone fractures and the anterior nasal spine.
Read definition & exam tips →
Sacrum and coccyx X-ray positioning (AP Axial & Lateral projections)
Radiographic positioning of the sacrum and coccyx requires opposite central ray (CR) angulations on the AP axial projections because of the posterior-convex pelvic curve. For the AP axial sacrum, the central ray is angled 15 degrees cephalad and centered 2 inches (5 cm) superior to the pubic symphysis along the midsagittal plane to project the sacrum free of foreshortening. For the AP axial coccyx, the central ray is angled 10 degrees caudad and centered 2 inches (5 cm) superior to the pubic symphysis so the coccygeal segments are projected superior to the pubic bones. For the lateral sacrum and coccyx projection, the perpendicular central ray is centered 3 to 4 inches (8 to 10 cm) posterior to the anterior superior iliac spine (ASIS), using a lead blocker behind the patient to absorb scatter radiation.
Read definition & exam tips →
Sternum X-ray positioning (15-20 degree RAO & Lateral projections)
Routine radiography of the sternum consists of a 15- to 20-degree right anterior oblique (RAO) projection and a lateral projection. A 15- to 20-degree RAO (rather than an LAO or direct PA) is used to project the sternum slightly to the left of the thoracic vertebrae and superimpose it over the homogeneous shadow of the heart rather than the air-filled right lung. A thin-chested (asthenic) patient requires a slightly larger obliquity (~20 degrees) than a deep-chested (hypersthenic) patient (~15 degrees). Most protocols use an orthostatic (shallow breathing) exposure technique (3- to 4-second exposure time with low mA) on the RAO sternum to blur overlying posterior rib markings, while the lateral sternum is acquired on suspended deep inspiration (with arms drawn posteriorly) to protrude the sternum anteriorly.
Read definition & exam tips →
Scapula X-ray positioning (AP & Lateral Y-view projections)
The routine scapula series includes an AP scapula projection and a lateral scapula (RAO or LAO) projection. For the AP scapula, the affected arm is gently abducted 90 degrees with the elbow flexed and the hand supinated (placed beside the head) to draw the lateral portion of the scapula clear of the thoracic rib cage; the perpendicular central ray enters 2 inches (5 cm) inferior to the coracoid process (mid-scapular area), using a slow breathing technique to blur overlying rib detail. For the lateral scapula, the patient is rotated 45 to 60 degrees into an RAO or LAO until the flat body of the scapula is perpendicular to the image receptor, placing the affected arm across the chest to evaluate the scapular body or hanging at the side to evaluate the acromion and coracoid processes.
Read definition & exam tips →