Radiography & X-Ray Physics (ARRT)

Milliamperage and mAs

Also called: mA, mAs, milliampere-seconds

Definition

Milliamperage (mA) measures the X-ray tube current, which determines the rate at which electrons flow from the cathode filament to the anode target. Multiplying tube current (mA) by exposure time in seconds (s) gives milliampere-seconds (mAs), the primary technical factor controlling the total quantity of X-ray photons produced and the patient radiation dose. Doubling mAs doubles both the number of photons reaching the image receptor and the patient dose, without changing the maximum energy or penetrating power of the beam.

How Milliamperage and mAs Is Tested on the Registry Exam

ARRT Radiography questions test mAs reciprocity (different mA and time combinations that yield the same total mAs), the direct 1-to-1 proportionality between mAs and receptor exposure or patient dose, and why mAs controls quantum noise rather than beam penetration.

ARRT Radiography Exam Overview (2026)

Milliamperage and mAs is part of the core curriculum for the ARRT Radiography Certification (R.T.(R)) exam.

Full Radiography Study Guide →
Questions
230 (200 scored)
Time Limit
3 hours 50 minutes
Passing Score
75 (1-99)
Test Provider
Pearson VUE

Official Radiography Content Domains

Patient Care16.5%
Safety25%
Image Production25.5%
Procedures33%

Video Lessons Covering Milliamperage and mAs

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

Related Radiography Physics, Image Production and Radiation Safety Terms

View all 22 terms →

Kilovoltage peak (kVp)

Kilovoltage peak (kVp) is the maximum electrical potential difference applied across the X-ray tube between the cathode and anode, measured in thousands of volts. It sets the maximum kinetic energy of the projectile electrons and therefore the maximum energy and penetrating ability (quality) of the resulting X-ray beam. Raising kVp increases both the average photon energy and the percentage of photons that penetrate the patient to reach the image receptor, while also increasing the proportion of Compton scatter.

Read definition & exam tips →

15 percent kVp rule

The 15 percent kVp rule is a practical radiographic exposure relationship stating that increasing kVp by 15 percent has the same effect on image receptor exposure as doubling the mAs, while decreasing kVp by 15 percent has the same effect as cutting the mAs in half. When a technologist wants to maintain equal receptor exposure while reducing patient radiation dose, raising kVp by 15 percent and halving mAs lowers the entrance skin dose because fewer total photons are needed when each photon is more penetrating.

Read definition & exam tips →

Dose-area product (DAP)

Dose-area product (DAP), also called kerma-area product (KAP), is the absorbed radiation dose in air multiplied by the cross-sectional area of the X-ray beam at that point, commonly expressed in milligray-centimeters squared (mGy·cm2) or microgray-meters squared. Unlike air kerma at a single point, DAP accounts for both the intensity of the beam and the field size (collimation). Because scattered radiation shifts proportionally with beam area, tightly collimating the X-ray field directly reduces DAP even if the central-ray exposure factors remain unchanged.

Read definition & exam tips →

Detective quantum efficiency (DQE)

Detective quantum efficiency (DQE) measures how effectively a digital image receptor converts incoming X-ray photons into a useful output image signal without adding extra noise, expressed as a percentage or ratio from 0 to 1. A receptor with a higher DQE can produce the same signal-to-noise ratio (image quality) at a lower patient radiation dose. Cesium iodide (CsI) indirect flat-panel detectors and amorphous selenium direct detectors generally achieve higher DQE values than computed radiography (CR) photostimulable phosphor plates.

Read definition & exam tips →

Inverse square law

The inverse square law states that the intensity of an X-ray beam is inversely proportional to the square of the distance from the point source. As the beam travels outward from the focal spot, the same number of photons spreads over an area that grows with the square of the distance. Doubling the distance from the source reduces radiation intensity to one-fourth of its original value, while halving the distance quadruples the intensity.

Read definition & exam tips →

Anode heel effect

The anode heel effect is the variation in X-ray beam intensity along the cathode-to-anode axis of the tube, caused by target material absorbing some of the photons emitted toward the anode side. As a result, the beam is more intense on the cathode side and up to 25 to 45 percent less intense on the anode side. The effect becomes more pronounced with smaller anode target angles, shorter source-to-image receptor distances (SID), and larger field sizes. Technologists use it clinically by placing the thicker, denser part of the anatomy under the cathode side.

Read definition & exam tips →