Bernoulli principle
Definition
The Bernoulli principle states that as a flowing fluid's velocity increases, the pressure it exerts decreases, so total energy stays roughly constant when frictional losses are small. Blood speeds up as it crosses a narrowed vessel segment or stenotic valve to maintain the same volume of flow, so pressure drops at that point and rises again just past it. A simplified version of the equation, pressure gradient equals 4 times velocity squared, converts a measured peak velocity into an estimated pressure gradient.
How Bernoulli principle Is Tested on the Registry Exam
Echo questions use the simplified Bernoulli equation, pressure gradient equals 4 times velocity squared, to estimate pressure gradients across stenotic or regurgitant heart valves from Doppler velocities.
ARDMS SPI Exam Overview (2026)
Bernoulli principle is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Textbook & Clinical References
Related Doppler and Hemodynamics Terms
View all 14 terms →Continuous wave Doppler
Continuous wave Doppler uses two separate piezoelectric elements, one continuously transmitting and one continuously receiving, so it can measure very high velocities accurately without a Nyquist limit or aliasing. Its major limitation is range ambiguity: because it samples along the entire beam at once rather than in pulses, it cannot localize the depth at which a given Doppler signal arose, giving it the poorest axial resolution of any Doppler mode. It still reports both flow speed and direction.
Read definition & exam tips →
Color Doppler
Color Doppler overlays a real-time, color-coded map of blood flow direction and mean velocity onto a grayscale image, using pulsed wave Doppler sampled at many points at once rather than the Fast Fourier Transform used for spectral display. By convention, flow toward the transducer is one color, typically red, and flow away is another, typically blue, with brighter shades showing higher mean velocity. It shares the same Nyquist limit and aliasing behavior as pulsed wave spectral Doppler.
Read definition & exam tips →
Doppler shift
Doppler shift is the difference between the frequency a transducer transmits and the frequency it receives back, caused by relative motion between the transducer and a moving reflector such as red blood cells. It is directly proportional to flow velocity and to transducer frequency, and is largest at a 0 degree angle of incidence and smallest near 90 degrees. Flow moving toward the transducer raises the received frequency, a positive shift, and flow moving away lowers it, a negative shift.
Read definition & exam tips →
Doppler angle
Doppler angle is the angle between the ultrasound beam and the direction of blood flow, and it should be kept at 60 degrees or less, and consistent between measurements, for an accurate velocity estimate. Because the Doppler equation includes the cosine of this angle, error increases sharply beyond 60 degrees, so small inconsistencies in angle correction produce large velocity errors; at 90 degrees, cosine equals zero and no shift is measured regardless of true flow. Tortuous vessels can make holding 60 degrees difficult.
Read definition & exam tips →
Nyquist limit
The Nyquist limit is the highest Doppler shift frequency that pulsed wave or color Doppler can sample and display correctly, equal to half the pulse repetition frequency. A true shift greater than this limit cannot be sampled often enough to display accurately and will alias. Because PRF is tied to depth, decreasing the depth of the sample volume raises PRF and therefore the Nyquist limit, one way to resolve aliasing at high velocities.
Read definition & exam tips →
Aliasing
Aliasing is a display error in pulsed wave and color Doppler where a true velocity exceeding the Nyquist limit is shown incorrectly, wrapping around to appear as flow in the opposite direction. On a spectral waveform the peaks above the limit appear cut off at the top and wrapped to the bottom; on color Doppler it appears as an abrupt switch from one flow color to its opposite. It can be corrected by raising PRF or lowering the baseline; continuous wave Doppler does not alias.
Read definition & exam tips →