Ultrasound Physics (ARDMS SPI)

Far field

Also called: Fraunhofer zone

Definition

The far field, or Fraunhofer zone, is the part of the beam beyond the focus, where it diverges and spreads with increasing depth. A small, low-frequency transducer spreads the most; a large, high-frequency one spreads the least, and less spread means better lateral resolution out here. Because the beam is wider here than at the focus, lateral resolution steadily worsens the farther a structure lies past the focal zone.

How Far field Is Tested on the Registry Exam

SPI questions test that a large-diameter, high-frequency transducer gives the least beam divergence, and therefore the best lateral resolution, in the far field.

ARDMS SPI Exam Overview (2026)

Far field is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Far field

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound. (p. 187)

Related Ultrasound Physics and Instrumentation Terms

View all 27 terms →

Time gain compensation

Time gain compensation, or TGC, also called depth gain compensation, amplifies returning echoes according to the depth they came from, boosting deeper echoes more than shallow ones to counteract attenuation. Without it, structures of equal true brightness would look progressively dimmer with depth simply because more sound was lost getting there and back. TGC lets the sonographer adjust this depth-dependent amplification to produce a uniformly bright image.

Read definition & exam tips →

Dynamic range

Dynamic range is the ratio between the largest and smallest echo amplitudes a system can process and display, expressed in decibels. A wide dynamic range shows many shades of gray, a softer-looking, lower-contrast image, while a narrow dynamic range shows fewer shades and higher contrast. Compression shrinks that ratio before display: apply 50 dB of compression to a 90 dB signal and 40 dB reaches the screen.

Read definition & exam tips →

Harmonic imaging

Harmonic imaging forms the image from harmonic frequencies generated as sound distorts while traveling through tissue, rather than from the reflected fundamental frequency. The second harmonic used is exactly twice the transmitted frequency. Because harmonic energy builds up gradually with depth and is weakest in the off-axis side lobes, tissue harmonic imaging reduces near-field reverberation clutter and lobe artifact compared with fundamental imaging, improving contrast resolution in technically difficult patients.

Read definition & exam tips →

Anechoic, hypoechoic, isoechoic, and hyperechoic

Echogenicity describes how bright a tissue or lesion appears on grayscale B-mode ultrasound relative to adjacent reference tissue. An anechoic structure produces zero internal echoes and displays completely black, which is the classic appearance of simple fluid such as a simple cyst, the urinary bladder, or a patent blood vessel lumen. Hypoechoic structures return weaker echoes and look darker gray than surrounding parenchyma, isoechoic structures match the brightness of the adjacent tissue, and hyperechoic structures return stronger echoes and appear brighter white (such as fat, fibrous tissue, calcifications, or air).

Read definition & exam tips →

Mechanical index and thermal index

Mechanical index (MI) and thermal index (TI) are the two real-time on-screen output display standard (ODS) bioeffect indices that help sonographers practice ALARA. Mechanical index estimates the likelihood of non-thermal mechanical bioeffects, specifically cavitation (microbubble expansion and collapse), and is proportional to peak rarefactional pressure divided by the square root of frequency, meaning lower frequency and higher acoustic output raise MI. Thermal index estimates the potential temperature rise in tissue (TIS for soft tissue, TIB for bone at the focus, and TIC for cranial bone at the surface), where a TI of 1.0 indicates a potential 1 degree Celsius rise.

Read definition & exam tips →

Coded excitation

Coded excitation is a transmit processing technique created in the pulser that sends a long, mathematically encoded sequence of pulses (such as a frequency-modulated chirp or Golay code) into the body and then mathematically compresses the returning echo on receive into a very short, high-amplitude spike. Because the energy is spread across a longer transmit burst rather than a single high-voltage spike, coded excitation increases signal-to-noise ratio, penetration, and contrast resolution while preserving axial resolution and keeping peak intensity (and mechanical index) within safe bioeffect limits.

Read definition & exam tips →