Ultrasound Physics (ARDMS SPI)

Piezoelectric effect

Definition

The piezoelectric effect is the property of certain crystals, such as lead zirconate titanate, that lets mechanical deformation from a returning echo generate an electrical signal on receive. Applying a voltage to the same crystal deforms it and produces sound on transmit, a process some textbooks call the reverse piezoelectric effect. Together these two conversions, electrical to mechanical and mechanical to electrical, let a single transducer both send and receive ultrasound.

How Piezoelectric effect Is Tested on the Registry Exam

SPI questions ask you to identify the piezoelectric effect as the mechanism converting electrical energy to sound on transmit and echoes back into an electrical signal on receive.

ARDMS SPI Exam Overview (2026)

Piezoelectric effect is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Piezoelectric effect

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound. (p. 113)

Related Ultrasound Physics and Instrumentation Terms

View all 27 terms →

Focal zone

The focal zone is the region of the beam surrounding the focus where it is narrowest, giving the best lateral resolution available at that depth, with roughly half the zone in the distal near field and half in the proximal far field. Sonographers place it at or just below the structure of interest to sharpen lateral resolution there. Using a single focal zone rather than several also keeps frame rate, and therefore temporal resolution, higher.

Read definition & exam tips →

Near field

The near field, or Fresnel zone, is the part of the beam between the transducer face and the focus, where the beam converges and narrows. Its length grows with a larger transducer diameter or a higher frequency, which lets the focus sit deeper and can improve resolution further from the skin. Structures too shallow for a clear image can be pushed into the near field's useful range with an acoustic standoff pad.

Read definition & exam tips →

Far field

The far field, or Fraunhofer zone, is the part of the beam beyond the focus, where it diverges and spreads with increasing depth. A small, low-frequency transducer spreads the most; a large, high-frequency one spreads the least, and less spread means better lateral resolution out here. Because the beam is wider here than at the focus, lateral resolution steadily worsens the farther a structure lies past the focal zone.

Read definition & exam tips →

Time gain compensation

Time gain compensation, or TGC, also called depth gain compensation, amplifies returning echoes according to the depth they came from, boosting deeper echoes more than shallow ones to counteract attenuation. Without it, structures of equal true brightness would look progressively dimmer with depth simply because more sound was lost getting there and back. TGC lets the sonographer adjust this depth-dependent amplification to produce a uniformly bright image.

Read definition & exam tips →

Dynamic range

Dynamic range is the ratio between the largest and smallest echo amplitudes a system can process and display, expressed in decibels. A wide dynamic range shows many shades of gray, a softer-looking, lower-contrast image, while a narrow dynamic range shows fewer shades and higher contrast. Compression shrinks that ratio before display: apply 50 dB of compression to a 90 dB signal and 40 dB reaches the screen.

Read definition & exam tips →

Harmonic imaging

Harmonic imaging forms the image from harmonic frequencies generated as sound distorts while traveling through tissue, rather than from the reflected fundamental frequency. The second harmonic used is exactly twice the transmitted frequency. Because harmonic energy builds up gradually with depth and is weakest in the off-axis side lobes, tissue harmonic imaging reduces near-field reverberation clutter and lobe artifact compared with fundamental imaging, improving contrast resolution in technically difficult patients.

Read definition & exam tips →