Near field
Also called: Fresnel zone
Definition
The near field, or Fresnel zone, is the part of the beam between the transducer face and the focus, where the beam converges and narrows. Its length grows with a larger transducer diameter or a higher frequency, which lets the focus sit deeper and can improve resolution further from the skin. Structures too shallow for a clear image can be pushed into the near field's useful range with an acoustic standoff pad.
How Near field Is Tested on the Registry Exam
SPI questions ask how increasing transducer diameter or frequency lengthens the near zone and moves the focus deeper.
ARDMS SPI Exam Overview (2026)
Near field is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Near field
Review these related video lessons from Prepry's registry review courses.
Textbook & Clinical References
- Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.
Related Ultrasound Physics and Instrumentation Terms
View all 27 terms →Far field
The far field, or Fraunhofer zone, is the part of the beam beyond the focus, where it diverges and spreads with increasing depth. A small, low-frequency transducer spreads the most; a large, high-frequency one spreads the least, and less spread means better lateral resolution out here. Because the beam is wider here than at the focus, lateral resolution steadily worsens the farther a structure lies past the focal zone.
Read definition & exam tips →
Time gain compensation
Time gain compensation, or TGC, also called depth gain compensation, amplifies returning echoes according to the depth they came from, boosting deeper echoes more than shallow ones to counteract attenuation. Without it, structures of equal true brightness would look progressively dimmer with depth simply because more sound was lost getting there and back. TGC lets the sonographer adjust this depth-dependent amplification to produce a uniformly bright image.
Read definition & exam tips →
Dynamic range
Dynamic range is the ratio between the largest and smallest echo amplitudes a system can process and display, expressed in decibels. A wide dynamic range shows many shades of gray, a softer-looking, lower-contrast image, while a narrow dynamic range shows fewer shades and higher contrast. Compression shrinks that ratio before display: apply 50 dB of compression to a 90 dB signal and 40 dB reaches the screen.
Read definition & exam tips →
Harmonic imaging
Harmonic imaging forms the image from harmonic frequencies generated as sound distorts while traveling through tissue, rather than from the reflected fundamental frequency. The second harmonic used is exactly twice the transmitted frequency. Because harmonic energy builds up gradually with depth and is weakest in the off-axis side lobes, tissue harmonic imaging reduces near-field reverberation clutter and lobe artifact compared with fundamental imaging, improving contrast resolution in technically difficult patients.
Read definition & exam tips →
Anechoic, hypoechoic, isoechoic, and hyperechoic
Echogenicity describes how bright a tissue or lesion appears on grayscale B-mode ultrasound relative to adjacent reference tissue. An anechoic structure produces zero internal echoes and displays completely black, which is the classic appearance of simple fluid such as a simple cyst, the urinary bladder, or a patent blood vessel lumen. Hypoechoic structures return weaker echoes and look darker gray than surrounding parenchyma, isoechoic structures match the brightness of the adjacent tissue, and hyperechoic structures return stronger echoes and appear brighter white (such as fat, fibrous tissue, calcifications, or air).
Read definition & exam tips →
Mechanical index and thermal index
Mechanical index (MI) and thermal index (TI) are the two real-time on-screen output display standard (ODS) bioeffect indices that help sonographers practice ALARA. Mechanical index estimates the likelihood of non-thermal mechanical bioeffects, specifically cavitation (microbubble expansion and collapse), and is proportional to peak rarefactional pressure divided by the square root of frequency, meaning lower frequency and higher acoustic output raise MI. Thermal index estimates the potential temperature rise in tissue (TIS for soft tissue, TIB for bone at the focus, and TIC for cranial bone at the surface), where a TI of 1.0 indicates a potential 1 degree Celsius rise.
Read definition & exam tips →