Doppler Ultrasound (SPI & RVT)

Pulsatility index

Also called: PI

Definition

Pulsatility index, or PI, is a Doppler measurement of flow resistance, calculated as peak systolic velocity minus end-diastolic velocity, divided by the mean velocity averaged over the cardiac cycle. A higher PI reflects greater downstream resistance, as seen when clenching a fist raises distal resistance in a radial artery waveform, while a lower PI reflects vasodilation, as in the fetal middle cerebral artery brain-sparing response to hypoxia. PI stays defined even with reversed diastolic flow, unlike resistive index.

How Pulsatility index Is Tested on the Registry Exam

OB and vascular questions use a falling middle cerebral artery PI as the sign of fetal brain-sparing vasodilation in growth restriction.

ARDMS SPI Exam Overview (2026)

Pulsatility index is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Pulsatility index

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

Related Doppler and Hemodynamics Terms

View all 14 terms →

Laminar flow

Laminar flow is the normal pattern of blood movement in a healthy, unobstructed vessel, with concentric layers moving in parallel, fastest at the center and progressively slower toward the vessel wall due to friction. It occurs when the Reynolds number, a unitless value based on flow speed, vessel diameter, density, and viscosity, stays below about 1,500. On a spectral Doppler waveform it produces a narrow band of velocities at each instant, leaving a clear window beneath the waveform.

Read definition & exam tips →

Turbulent flow

Turbulent flow is a disorganized pattern in which blood moves in multiple directions and at many velocities at the same location and instant, rather than in smooth parallel layers. It develops when the Reynolds number rises above roughly 2,000, commonly distal to a stenosis, vessel wall irregularity, or high-velocity jet. On a spectral waveform it fills in the window as spectral broadening, and on color Doppler it can appear as a mosaic of mixed colors within the lumen.

Read definition & exam tips →

Bernoulli principle

The Bernoulli principle states that as a flowing fluid's velocity increases, the pressure it exerts decreases, so total energy stays roughly constant when frictional losses are small. Blood speeds up as it crosses a narrowed vessel segment or stenotic valve to maintain the same volume of flow, so pressure drops at that point and rises again just past it. A simplified version of the equation, pressure gradient equals 4 times velocity squared, converts a measured peak velocity into an estimated pressure gradient.

Read definition & exam tips →

Continuous wave Doppler

Continuous wave Doppler uses two separate piezoelectric elements, one continuously transmitting and one continuously receiving, so it can measure very high velocities accurately without a Nyquist limit or aliasing. Its major limitation is range ambiguity: because it samples along the entire beam at once rather than in pulses, it cannot localize the depth at which a given Doppler signal arose, giving it the poorest axial resolution of any Doppler mode. It still reports both flow speed and direction.

Read definition & exam tips →

Color Doppler

Color Doppler overlays a real-time, color-coded map of blood flow direction and mean velocity onto a grayscale image, using pulsed wave Doppler sampled at many points at once rather than the Fast Fourier Transform used for spectral display. By convention, flow toward the transducer is one color, typically red, and flow away is another, typically blue, with brighter shades showing higher mean velocity. It shares the same Nyquist limit and aliasing behavior as pulsed wave spectral Doppler.

Read definition & exam tips →

Doppler shift

Doppler shift is the difference between the frequency a transducer transmits and the frequency it receives back, caused by relative motion between the transducer and a moving reflector such as red blood cells. It is directly proportional to flow velocity and to transducer frequency, and is largest at a 0 degree angle of incidence and smallest near 90 degrees. Flow moving toward the transducer raises the received frequency, a positive shift, and flow moving away lowers it, a negative shift.

Read definition & exam tips →