Ultrasound Physics (ARDMS SPI)

Temporal resolution

Definition

Temporal resolution is how accurately a system displays motion over time, and in practice it is the frame rate: faster frame rates give better temporal resolution. Frame rate rises when pulse repetition frequency increases and falls when the system must acquire more information per frame, so increasing imaging depth, sector width, line density, or the number of focal zones all lower the frame rate and worsen temporal resolution.

How Temporal resolution Is Tested on the Registry Exam

SPI questions ask what happens to frame rate, and therefore temporal resolution, when depth, sector width, line density, or focal zones increase.

ARDMS SPI Exam Overview (2026)

Temporal resolution is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Temporal resolution

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.

Related Ultrasound Physics and Instrumentation Terms

View all 27 terms →

Frequency

Frequency is the number of complete sound wave cycles per second, measured in Hertz, and diagnostic ultrasound uses roughly 2 to 18 MHz. It is set by the transducer and stays constant as sound crosses into a new medium, unlike wavelength, which changes with propagation speed. Higher frequency shortens wavelength and improves resolution, but it also raises scattering and attenuation, which is why higher frequencies penetrate less deeply.

Read definition & exam tips →

Wavelength

Wavelength is the physical length that one complete cycle of a sound wave occupies, equal to propagation speed divided by frequency. Because frequency is fixed by the transducer, wavelength changes only when propagation speed changes, such as when sound crosses into a different medium; a faster medium stretches the wavelength and a slower one shortens it. In soft tissue, wavelength in millimeters is roughly 1.54 divided by frequency in MHz.

Read definition & exam tips →

Period

Period is the time required to complete one full cycle of a sound wave, usually measured in microseconds, and it equals one divided by frequency. Like frequency, period is fixed by the sound source and does not change when the wave enters a different medium, even though wavelength does. A 5 MHz transducer, for example, produces a period of 0.2 microseconds. Pulse repetition frequency and pulse repetition period are separate, related quantities describing how often pulses are sent.

Read definition & exam tips →

Pulse repetition frequency

Pulse repetition frequency, or PRF, is the number of ultrasound pulses the transducer sends out each second, and it has a reciprocal relationship with pulse repetition period. PRF is largely set by imaging depth, since the system must wait for echoes to return from the deepest structure before firing the next pulse, so deeper imaging requires a lower PRF. PRF also sets the Nyquist limit, equal to half of PRF, and pushing PRF very high to raise that limit can cause range ambiguity.

Read definition & exam tips →

Pulse repetition period

Pulse repetition period, or PRP, is the time from the start of one pulse to the start of the next, and it equals one divided by pulse repetition frequency. PRP is made up of pulse duration, the brief transmit time, plus the much longer listening time spent waiting for echoes to return. Increasing imaging depth lengthens PRP because the system must wait longer for the deepest echoes, while decreasing depth shortens it.

Read definition & exam tips →

Spatial pulse length

Spatial pulse length, or SPL, is the physical length of space that one ultrasound pulse occupies, equal to the number of cycles in the pulse multiplied by wavelength. It is set entirely by the transducer and the medium, not by the sonographer or imaging depth, and typically ranges from 0.1 to 1.0 mm clinically. SPL is the only variable that directly sets axial resolution, since axial resolution equals SPL divided by two, so fewer cycles or a shorter wavelength both shorten it.

Read definition & exam tips →