Ultrasound Physics (ARDMS SPI)

Pulse repetition period

Definition

Pulse repetition period, or PRP, is the time from the start of one pulse to the start of the next, and it equals one divided by pulse repetition frequency. PRP is made up of pulse duration, the brief transmit time, plus the much longer listening time spent waiting for echoes to return. Increasing imaging depth lengthens PRP because the system must wait longer for the deepest echoes, while decreasing depth shortens it.

How Pulse repetition period Is Tested on the Registry Exam

SPI questions ask you to calculate PRP from a given PRF, or to identify how imaging depth changes PRP.

ARDMS SPI Exam Overview (2026)

Pulse repetition period is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Pulse repetition period

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.

Related Ultrasound Physics and Instrumentation Terms

View all 27 terms →

Spatial pulse length

Spatial pulse length, or SPL, is the physical length of space that one ultrasound pulse occupies, equal to the number of cycles in the pulse multiplied by wavelength. It is set entirely by the transducer and the medium, not by the sonographer or imaging depth, and typically ranges from 0.1 to 1.0 mm clinically. SPL is the only variable that directly sets axial resolution, since axial resolution equals SPL divided by two, so fewer cycles or a shorter wavelength both shorten it.

Read definition & exam tips →

Duty factor

Duty factor is the fraction of time the transducer spends actively transmitting sound rather than listening for echoes, equal to pulse duration divided by pulse repetition period, expressed as a percentage. Pulsed imaging has a duty factor under 1 percent because most of each cycle is spent listening, while continuous wave Doppler transmits constantly and has a duty factor of 100 percent. Increasing imaging depth lengthens listening time and lowers duty factor.

Read definition & exam tips →

Damping

Damping is the reduction of ringing in a piezoelectric element after it is excited, done with a backing material bonded to the back of the crystal that absorbs excess vibration. Heavier damping shortens the pulse and spatial pulse length, which improves axial resolution and widens bandwidth, at the cost of reduced sensitivity. A detached or failing backing layer is a recognized cause of an enlarged dead zone on quality assurance testing.

Read definition & exam tips →

Q factor

Q factor, or quality factor, describes how pure a range of frequencies a transducer's pulse contains, and it equals operating frequency divided by bandwidth, so it is inversely related to bandwidth. A high Q factor means a long, narrow-bandwidth pulse, typical of an undamped continuous wave transducer. Diagnostic imaging transducers use heavy backing material to lower Q factor, which widens bandwidth and shortens the pulse for better axial resolution.

Read definition & exam tips →

Bandwidth

Bandwidth is the spread of frequencies a pulse carries instead of one pure tone, measured in Hertz as the difference between the pulse's highest and lowest frequency. Shorter, heavily damped pulses contain a wider range of frequencies and so have wider bandwidth, which goes along with a short spatial pulse length and good axial resolution. Continuous wave transducers, by contrast, have a narrow bandwidth and a high Q factor.

Read definition & exam tips →

Piezoelectric effect

The piezoelectric effect is the property of certain crystals, such as lead zirconate titanate, that lets mechanical deformation from a returning echo generate an electrical signal on receive. Applying a voltage to the same crystal deforms it and produces sound on transmit, a process some textbooks call the reverse piezoelectric effect. Together these two conversions, electrical to mechanical and mechanical to electrical, let a single transducer both send and receive ultrasound.

Read definition & exam tips →