Period

Definition

Period is the time required to complete one full cycle of a sound wave, usually measured in microseconds, and it equals one divided by frequency. Like frequency, period is fixed by the sound source and does not change when the wave enters a different medium, even though wavelength does. A 5 MHz transducer, for example, produces a period of 0.2 microseconds. Pulse repetition frequency and pulse repetition period are separate, related quantities describing how often pulses are sent.

How Period Is Tested on the Registry Exam

SPI questions test which quantities, like period and frequency, stay constant across media, unlike wavelength, which changes with propagation speed.

ARDMS SPI Exam Overview (2026)

Period is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.

Full SPI Study Guide →
Questions
110
Time Limit
2 hours
Passing Score
555 (300-700)
Test Provider
Pearson VUE

Official SPI Content Domains

Perform Ultrasound Examinations23%
Manage Ultrasound Transducers7%
Optimize Sonographic Images26%
Apply Doppler Concepts34%
Provide Clinical Safety & Quality Assurance10%

Video Lessons Covering Period

Review these related video lessons from Prepry's registry review courses.

Textbook & Clinical References

  • Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.

Related Ultrasound Physics and Instrumentation Terms

View all 27 terms →

Pulse repetition frequency

Pulse repetition frequency, or PRF, is the number of ultrasound pulses the transducer sends out each second, and it has a reciprocal relationship with pulse repetition period. PRF is largely set by imaging depth, since the system must wait for echoes to return from the deepest structure before firing the next pulse, so deeper imaging requires a lower PRF. PRF also sets the Nyquist limit, equal to half of PRF, and pushing PRF very high to raise that limit can cause range ambiguity.

Read definition & exam tips →

Pulse repetition period

Pulse repetition period, or PRP, is the time from the start of one pulse to the start of the next, and it equals one divided by pulse repetition frequency. PRP is made up of pulse duration, the brief transmit time, plus the much longer listening time spent waiting for echoes to return. Increasing imaging depth lengthens PRP because the system must wait longer for the deepest echoes, while decreasing depth shortens it.

Read definition & exam tips →

Spatial pulse length

Spatial pulse length, or SPL, is the physical length of space that one ultrasound pulse occupies, equal to the number of cycles in the pulse multiplied by wavelength. It is set entirely by the transducer and the medium, not by the sonographer or imaging depth, and typically ranges from 0.1 to 1.0 mm clinically. SPL is the only variable that directly sets axial resolution, since axial resolution equals SPL divided by two, so fewer cycles or a shorter wavelength both shorten it.

Read definition & exam tips →

Duty factor

Duty factor is the fraction of time the transducer spends actively transmitting sound rather than listening for echoes, equal to pulse duration divided by pulse repetition period, expressed as a percentage. Pulsed imaging has a duty factor under 1 percent because most of each cycle is spent listening, while continuous wave Doppler transmits constantly and has a duty factor of 100 percent. Increasing imaging depth lengthens listening time and lowers duty factor.

Read definition & exam tips →

Damping

Damping is the reduction of ringing in a piezoelectric element after it is excited, done with a backing material bonded to the back of the crystal that absorbs excess vibration. Heavier damping shortens the pulse and spatial pulse length, which improves axial resolution and widens bandwidth, at the cost of reduced sensitivity. A detached or failing backing layer is a recognized cause of an enlarged dead zone on quality assurance testing.

Read definition & exam tips →

Q factor

Q factor, or quality factor, describes how pure a range of frequencies a transducer's pulse contains, and it equals operating frequency divided by bandwidth, so it is inversely related to bandwidth. A high Q factor means a long, narrow-bandwidth pulse, typical of an undamped continuous wave transducer. Diagnostic imaging transducers use heavy backing material to lower Q factor, which widens bandwidth and shortens the pulse for better axial resolution.

Read definition & exam tips →