Elevational resolution
Definition
Elevational resolution, also called slice-thickness resolution, is how well the system separates structures across the slice, the dimension at right angles to the scan plane, and it depends on how thick the beam is rather than how wide or how short the pulse is. It is fixed by the transducer's acoustic lens or elevation focus and cannot be adjusted at the console. A beam thicker than the structure being imaged pulls in echoes from outside the true plane, which can make a simple cyst appear to contain debris.
How Elevational resolution Is Tested on the Registry Exam
SPI questions test that poor elevational resolution can make an anechoic structure like a cyst falsely appear to contain internal echoes.
ARDMS SPI Exam Overview (2026)
Elevational resolution is part of the core curriculum for the Sonography Principles & Instrumentation (Required for RDMS/RDCS/RVT) exam.
- Questions
- 110
- Time Limit
- 2 hours
- Passing Score
- 555 (300-700)
- Test Provider
- Pearson VUE
Official SPI Content Domains
Video Lessons Covering Elevational resolution
Review these related video lessons from Prepry's registry review courses.
Textbook & Clinical References
- Edelman, S. K. (2012). Understanding Ultrasound Physics. ESP Ultrasound.
Related Ultrasound Physics and Instrumentation Terms
View all 27 terms →Temporal resolution
Temporal resolution is how accurately a system displays motion over time, and in practice it is the frame rate: faster frame rates give better temporal resolution. Frame rate rises when pulse repetition frequency increases and falls when the system must acquire more information per frame, so increasing imaging depth, sector width, line density, or the number of focal zones all lower the frame rate and worsen temporal resolution.
Read definition & exam tips →
Frequency
Frequency is the number of complete sound wave cycles per second, measured in Hertz, and diagnostic ultrasound uses roughly 2 to 18 MHz. It is set by the transducer and stays constant as sound crosses into a new medium, unlike wavelength, which changes with propagation speed. Higher frequency shortens wavelength and improves resolution, but it also raises scattering and attenuation, which is why higher frequencies penetrate less deeply.
Read definition & exam tips →
Wavelength
Wavelength is the physical length that one complete cycle of a sound wave occupies, equal to propagation speed divided by frequency. Because frequency is fixed by the transducer, wavelength changes only when propagation speed changes, such as when sound crosses into a different medium; a faster medium stretches the wavelength and a slower one shortens it. In soft tissue, wavelength in millimeters is roughly 1.54 divided by frequency in MHz.
Read definition & exam tips →
Period
Period is the time required to complete one full cycle of a sound wave, usually measured in microseconds, and it equals one divided by frequency. Like frequency, period is fixed by the sound source and does not change when the wave enters a different medium, even though wavelength does. A 5 MHz transducer, for example, produces a period of 0.2 microseconds. Pulse repetition frequency and pulse repetition period are separate, related quantities describing how often pulses are sent.
Read definition & exam tips →
Pulse repetition frequency
Pulse repetition frequency, or PRF, is the number of ultrasound pulses the transducer sends out each second, and it has a reciprocal relationship with pulse repetition period. PRF is largely set by imaging depth, since the system must wait for echoes to return from the deepest structure before firing the next pulse, so deeper imaging requires a lower PRF. PRF also sets the Nyquist limit, equal to half of PRF, and pushing PRF very high to raise that limit can cause range ambiguity.
Read definition & exam tips →
Pulse repetition period
Pulse repetition period, or PRP, is the time from the start of one pulse to the start of the next, and it equals one divided by pulse repetition frequency. PRP is made up of pulse duration, the brief transmit time, plus the much longer listening time spent waiting for echoes to return. Increasing imaging depth lengthens PRP because the system must wait longer for the deepest echoes, while decreasing depth shortens it.
Read definition & exam tips →